NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
Microprocessor-based cardiotachometerInstrument operates reliably even with stress-test electrocardiogram (ECG) signals subject to noise, baseline wandering, and amplitude change. It records heart rate from preamplified, single-lead ECG input signal and produces digital and analog heart-rate outputs which are fed elsewhere. Analog hardware processes ECG input signal, producing 10-ms pulse for each heartbeat. Microprocessor analyzes resulting pulse train, identifying irregular heartbeats and maintaining stable output during lead switching. Easily modified computer program provides analysis.
Document ID
19800000501
Acquisition Source
Legacy CDMS
Document Type
Other - NASA Tech Brief
Authors
Crosier, W. G.
(Technology Inc.)
Donaldson, J. A.
(Technology Inc.)
Date Acquired
August 10, 2013
Publication Date
April 1, 1981
Publication Information
Publication: NASA Tech Briefs
Volume: 5
Issue: 4
ISSN: 0145-319X
Subject Category
Life Sciences
Report/Patent Number
MSC-18775
Accession Number
80B10501
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.

Available Downloads

There are no available downloads for this record.
No Preview Available