Measurements of induced radioactivity in some LDEF samplesTwenty-six stainless steel trunnion samples, five aluminum end support retainer plate samples, two aluminum keel plate samples, and two titanium clips were analyzed. The shielded high purity germanium detectors used had efficiencies of 33, 54, and 80 pcts. at 1332 keV. Detector efficiencies as a function of energy and corrections for self absorption in the samples were determined with calibrated sources, unactivated control samples, and calculations. Several measurements were made on most samples. In the trunnion samples, Mn-54 and Co-57 were seen and limits were obtained for other isotopes. The results agree well with 1-D activation calculations for an anisotropic trapped proton model. In the aluminum samples, Na-22 and Be-7 were seen. Other results are presented.