Quantification of urinary uric acid in the presence of thymol and thimerosal by high-performance liquid chromatographyA high-performance liquid chromatographic method was developed as an alternative to automated enzymatic analysis of uric acid in human urine preserved with thymol and/or thimerosal. Uric acid (tR = 10 min) and creatinine (tR = 5 min) were separated and quantified during isocratic elution (0.025 M acetate buffer, pH 4.5) from a mu Bondapak C18 column. The uric-acid peak was identified chemically by incubating urine samples with uricase. The thymol/thimerosal peak appeared at 31 min during the washing step and did not interfere with the analysis. We validated the high-performance liquid chromatographic method for linearity, precision and accuracy, and the results were found to be excellent.
Document ID
20040173022
Acquisition Source
Legacy CDMS
Document Type
Reprint (Version printed in journal)
Authors
Chen, Y.
(KRUG Life Sciences Inc. Houston, TX 77058, United States)
Pietrzyk, R. A.
Whitson, P. A.
Date Acquired
August 22, 2013
Publication Date
February 28, 1997
Publication Information
Publication: Journal of chromatography. A
Volume: 763
Issue: 1-2
Subject Category
Life Sciences (General)
Distribution Limits
Public
Copyright
Other
Keywords
NASA Discipline Regulatory Physiology
NASA Center JSC

Available Downloads

There are no available downloads for this record.
No Preview Available