A microprocessor-based cardiotachometerThe development of a highly accurate and reliable cardiotachometer for measuring the heart rate of test subjects is discussed. It measures heart rate over the range of 30 to 250 beats/minute and gives instantaneous (beat to beat) updates on the system output so that occasional noise artifacts or ectopic beats could be more easily identified except that occasional missed beats caused by switching ECG leads should not cause a change in the output. The cardiotachometer uses an improved analog filter and R-wave detector and an Intel 8080A microprocessor to handle all of the logic and arithmetic necessary. By using the microprocessor, future hardware modifications could easily be made if functional changes were needed.